| Cat. No. | Range (μL) | Increment (μL) | Measured Volume (μL) | Accuracy | Precision | ||
|---|---|---|---|---|---|---|---|
| % | μL | % | μL | ||||
| SPA200125 | 0.1–2.5 | 0.01 | 2.5 | 2.5 | 0.0625 | 1.6 | 0.04 |
| 1.25 | 3 | 0.0375 | 3 | 0.0375 | |||
| 0.25 | 12 | 0.03 | 6 | 0.015 | |||
| SPA200510 | 0.5–10 | 0.1 | 10 | 1 | 0.1 | 0.8 | 0.08 |
| 5 | 2 | 0.1 | 1 | 0.05 | |||
| 1 | 2.5 | 0.025 | 1.5 | 0.015 | |||
| SPA200220 | 2–20 | 0.5 | 20 | 0.9 | 0.18 | 0.4 | 0.08 |
| 10 | 1.5 | 0.15 | 1 | 0.1 | |||
| 2 | 3 | 0.06 | 2 | 0.04 | |||
| SPA200550 | 5–50 | 1 | 50 | 0.6 | 0.3 | 0.3 | 0.15 |
| 25 | 0.8 | 0.2 | 0.4 | 0.1 | |||
| 5 | 2 | 0.1 | 2 | 0.1 | |||
| SPA210100 | 10–100 | 0.5 | 100 | 0.8 | 0.8 | 0.15 | 0.15 |
| 50 | 1 | 0.5 | 0.5 | 0.25 | |||
| 10 | 3 | 0.3 | 1.5 | 0.15 | |||
| SPA220200 | 20–200 | 1 | 200 | 0.6 | 1.2 | 0.15 | 0.3 |
| 100 | 0.7 | 0.7 | 0.3 | 0.3 | |||
| 20 | 2 | 0.4 | 0.8 | 0.16 | |||
| SPA211000 | 100–1000 | 5 | 1000 | 0.6 | 6 | 0.2 | 2 |
| 500 | 1 | 5 | 0.4 | 2 | |||
| 100 | 2 | 2 | 0.7 | 0.7 | |||
| Cat. No. | Range (μL) | Increment | Measured Volume (μL) | Accuracy | Precision | ||
|---|---|---|---|---|---|---|---|
| % | μL | % | μL | ||||
| SPA100005 | 5 | — | 5 | 1.3 | 0.065 | 1.2 | 0.06 |
| SPA100010 | 10 | — | 10 | 0.8 | 0.08 | 0.8 | 0.08 |
| SPA100050 | 50 | — | 50 | 0.5 | 0.25 | 0.3 | 0.15 |
| SPA100100 | 100 | — | 100 | 0.5 | 0.5 | 0.3 | 0.3 |
| SPA100500 | 500 | — | 500 | 0.3 | 1.5 | 0.2 | 1 |
| SPA101000 | 1000 | — | 1000 | 0.3 | 3 | 0.15 | 1.5 |